Table 24 Methods (continued)
Parameter Description Basic mode Expert mode
Curve filter Contact technical support before changing. Number of points used
for building the titration curve
x
Curvature filter Contact technical support before changing. Number of points used
for the curvature calculation
x
Detection threshold Contact technical support before changing. The minimum value of
the curvature for equivalence detection
x
Linearity threshold Contact technical support before changing. The minimum value of
the curvature used in the algorithm for equivalence calculation
x
EQP min. ordinate Minimum permitted value for this equivalence point x
EQP max. ordinate Maximum permitted value for this equivalence point x
Stop on flat signal Stops titration when a flat signal is detected x
Derivative filter Contact technical support before changing. Number of points used
for building the derivative used in flat signal detection
x
Min. derivative Contact technical support before changing. Minimum threshold on
derivative for flat signal detection
x
Max. derivative Contact technical support before changing. Maximum threshold on
derivative for flat signal detection
x
Derivative number Contact technical support before changing. Number of derivative
values between the minimum and maximum thresholds to stop the
titration on a flat signal
x
Titration calculation
Result x (Rx) name Name of the result x x
Rx max. resolution Maximum number of decimals to show for the result x
Rx hide The result is not shown if Yes is selected x x
Rx limit check Activate the sample result boundary check x
Rx min. Minimum permitted result value x
Rx max. Maximum permitted result value x
Rx QC min. Minimum permitted QC result value x
Rx QC max. Maximum permitted QC result value x
Rx unit Unit of the result x
Rx EQP index Index of the equivalence point used for the result calculation x
Rx molar weight Molar weight of analyte used for the result calculation x
Rx sample equivalents Number of exchanged equivalents of the sample (H+; e-; ...) x
Rx titrant equivalents Number of exchanged equivalents of the titrant (H+; e-; ...) x
Titration calculation using an equation
Result x (Rx) name Name of the result calculated with the equation x x
Rx max. resolution Maximum number of decimals to show for the result x
Rx hide The result is not shown if Yes is selected x x
Rx limit check Activate the sample result boundary check x
Rx min. Minimum permitted result value x
Change application parameters
70