Option Description
Options Sets the following options:
Set the application parameters view to expert or basic mode
When the instrument is set to off, sets the syringe to empty into the
titrant bottle
Changes the temperature display unit
Prints the measurement and derivative curves if a printer is connected
Specify if a balance is connected
Specify if an external propeller stirrer is connected.
Note: When the instrument is set to on, the application parameters view is
always set to basic mode. Changes made to application data with the
instrument in expert mode are kept.
Security Change the password. Specify which of these options are password
protected:
Settings (all items except Info)
Maintenance schedule and summary
Delete data log
Export data log
Electrode calibration
Titrant calibration
Sample volume calibration
Syringe verification test
3. Push Back.
Configure the instrument
76

Table of Contents