Message Solution
Send data
Connect a receiving device. Examine the device connections. Set the Auto Send setting to off. Refer to
Configure the measurement settings on page 11.
Add sample IDs from list
No valid data found No sample ID file was found on the USB flash drive.
Not able to read sampling
date.
Make sure that the date and time format is dd.mm.yyyy hh:mm.
The instrument cannot read
the Sample ID
Examine the text strings. Refer to the expanded user manual on the
manufacturer's website.
Problem/Error: Incorrect
date
Possible cause: The wrong
date format.
Make sure that the date and time format is dd.mm.yyyy hh:mm.
The sample ID list full. Data
has not been added.
Remove the sample IDs that are not used. Add a new sample ID.