Specifications
Specifications are subject to change without notice.
Specification Details
Measurement method Nephelometric
Regulatory Meets EPA Method 180.1
ASTM D7315 - Standard Test Method for Determination of Turbidity Above 1 Turbidity
Unit (TU) in Static Mode
ASTM D6655 - Standard Test Method for Determination of Turbidity Below 5 NTU in
Static Mode
Dimensions (W x D x H) 39.5 x 30.5 x 15.3 cm (15.6 x 12.0 x 6.02 in.)
Weight 3.0 kg (6.6 lb)
Enclosure IP30; indoor use only
Protection Class External power supply: Protection Class I; instrument: Protection Class II
Pollution degree 2
Installation category External power supply: Category II; instrument: Category I
Power requirements Instrument: 12 VDC, 3.4 A; power supply: 100–240 VAC, 50/60 Hz
Operating temperature 0 to 40 °C (32 to 104 °F)
Storage temperature –20 to 60 °C (–4 to 140 °F)
Humidity 5 to 95% relative humidity, non-condensing
Display 17.8 mm (7 in.) color touch screen
Light source Tungsten filament lamp
Measurement units NTU, EBC, Abs (absorbance), %T (% transmittance) and mg/L (degree)
Range NTU (Ratio on): 0–10,000 auto decimal
NTU (Ratio off): 0–40
EBC (Ratio on): 0–2450 auto decimal
EBC (Ratio off): 0–9.8
Absorbance
1
(auto range): 0–1.0
Transmittance
1
(%): 1.0–100
Degree (mg/L): 1–100
1
A filter assembly is necessary for absorbance or transmittance measurements
English 3