Lesion size and shape is determined by numerous factors including:
• probe active tip length
• impedance of tissue
• current intensity
• length of time that energy is applied
• water ow rate
• water temperature
Ensure proper slection of the appropriate sized active electrode tip to achieve the desired lesion size.
Active Tip size
Lesion Size and Shape
(T–60°C)
Typical Anatomy
Placement
Distal Projection,
Average
2 mm
4 – 6 mm, Oblate
Spheriod
Cervical Spine 52%
4 mm
10 – 12 mm,
Spherical
Lumbar Spine,
Knee, Hip
46%
5.5 mm 12 mm Spherical Thoracic Spine 47%
Lesion
Diameter
from Table
40% Volumetric
Distal Projection
Active
Tip