28
value (xx):
xx=06 EC range
xx=07 TDS range
xx=08 NaCl range
The instrument sends the “ACK” (6) character every time a command is
recognized and a “NAK” (21) character for invalid commands.
COMMANDS REQUIRING AN ANSWER
RAS Causes the instrument to send a complete set of readings in
according with the current range:
Conductivity and temperature reading on EC range.
TDS and temperature reading on TDS range.
NaCl and temperature reading on NaCl range.
The answer string contains:
Meter mode (2 chars):
06 - EC range
07 - TDS range
08 - NaCl range
Meter status (2 chars of status byte): represents a 8
bit hexadecimal encoding.
0x40 - TDS unit (0=ppm, 1=g/L)
0x20 - EC unit (0=µS, 1=mS)
0x10 - temperature probe is connected
0x01 - new GLP data available
0x02 - new SETUP parameter
Reading status (1 char): R - in range, O - over
range, U - under range.
The reading (corresponding to the selected range) -
7 ASCII chars, including sign and decimal point.
Temperature reading - 7 ASCII chars, with sign and
two decimal points, always in ºC.
MDR Requests the instrument model name and firmware code.
GLP Requests the calibration data record.
The answer string contains:
GLP status (1 char): represents a 4 bit hexadecimal
encoding.
0x04 - EC calibration available
0x08 - NaCl calibration available
sale@ponpe.com
www.ponpe.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 2300 and is the answer not in the manual?

Hanna Instruments HI 2300 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 2300
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals