SPECiFiCATiONS
pH Calibration
Calibration points Up to 5
Type
Automatic
Semiautomatic
Manual
Standard buffers
Hanna and NIST
pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45
Custom buffers Up to 5
Custom group Up to 5
1
st
calibration point Offset or Points (user setting)
Reminder
Disabled
Daily: 0 min. to 23 hours and 59 min.
Periodic: 1 min. to 500 days, 23 hours and 59 min.
pH Views
Basic
Measurement (pH, mV, Rel.mV, Abs.mV)
Temperature, Stability status
Simple GLP
Basic view information
Last calibration date, electrode offset, average
slope, and electrode condition
Full GLP Simple GLP information and calibration point details
Table
Measurements updated every second are displayed
in table
Graph (Plot)
pH (or mV) and temperature versus time graph can
be panned or zoomed (pinch-to-zoom technology)
Temperature Compensation Automatic or Manual
Logging
Type Automatic, Manual, Autohold
Number of records
50 000 maximum per file
Stores at least 1 000 000 data points per user
Automatic interval 1, 2, 5, 10, 30 seconds
1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes
Sample ID Incremental mode
Export option .csv file format
Users Up to 9 users and admin. account (default)