68
LOGGiNG
6. Tap Home key ( ) on instrument, to return to Measurement screen.
7. From Measurement screen, tap manual logging symbol (
1
) to log data.
Data is logged every time symbol is tapped.
8. See option to track and log notes.
Follow Automatic Logging, step 7.
14.3. AUTOHOLD LOGGING
1. From Measurement screen, tap Measurement Menu icon ( ).
2. Tap Reading tab to select stability criteria (Accurate, Medium, or Fast).
Note: Autohold logging uses this criteria for logging. Setting this will affect when data is recorded.
3. Tap for direct selection of Direct/Autohold reading mode.
4. Tap Logging tab and select Autohold logging type
5. Ceate a file name or have a file previously set up
6. Option to define sample ID (name and prefix)
• Tap Sample ID Prefix field to start editing.
• Use the on-screen keypad to enter lot file name.
• Tap Enter on keypad to save new name.
• Next, scroll to select Prefix value.
7. Tap Home key ( ) on instrument, to reenter Measurement screen.
8. From Measurement screen, tap Direct logging symbol (
D
) on display to start logging.
Active Autohold logging in progress is indicated by displayed
A
symbol.
Progress bar, top left corner, indicates progress status.
9. Once configured stability criteria is checked, and conditions are met, data is logged.