58
CALiBrATiON
5. For standard values, tap active forward arrow to move buffer in the Buffers in Use column.
A small s icon (
) on display, next to the buffer value, indicates calibration with standard
buffer. Repeat with up to 5 buffers.
6. For custom values, tap an empty tray and follow the Editing Custom Buffers Values steps.
Editing Custom Buffer Values (navigation guidelines)
With Calibration tab selected:
1. Tap Edit next to Buffer Group.
2. Tap to select an empty tray from the Available Custom Buffers column.
3. Tap Edit.
4. Use the numeric keypad to enter value.
5. Tap Enter on keypad to confirm (tap Cancel or Clear to return to editing mode).
6. Tap on tray with newly-entered custom value.
A rectangular outline confirms that buffer can be transferred to Buffers in Use.
7. Use forward arrow to move custom buffer into Buffers in Use column.