SETUP MODE
Setup mode allows the selection of the Temperature unit,
Auto‑off, Beep, the type of pH buffer set, the pH Resolu‑
tion and enable/disable calibration Information display. To
enter Setup mode press and hold button until “STD“
and tag is replaced by “SETUP“ and tag.
Release the button.
• “TEMP“ is displayed on the third
LCD line with the current temperature
unit (E.g. “TEMP °C“), for °C/°F
selection, use the SET button. After
the temperature unit has been selected, press to confirm
and to enter the “A-OFF“ selection.
• Use the SET button, to cycle through
the auto‑off choices: 8 minutes (“8”,
default value), 60 minutes (“60”) or
disabled (“---”). Press to confirm
and to enter the “BEEP“ selection.
• To switch ON or OFF the beep tone,
press the SET button; press to
confirm and to enter the calibration
buffer selection “pH 7.01 BUFF“.
• To change the type of calibration
buffer set, the meter will show the
current buffer set: “pH 7.01 BUFF”
(for standard buffer set: 4.01/7.01/
10.01) or “pH 6.86 BUFF” (for
NIST buffer set: 4.01/6.86/ 9.18). Change the set with the
SET button. Press to confirm and to enter pH resolution
selection “RESOL“.
• To change the pH measurement
resolution between “0.1“ and “0.01“
use the SET button; then press
to confirm and to enter electrode
calibration information “INFO“ selection.