SETUP MODE
Setup mode allows the selection of the Temperature unit,
Auto‑off, Beep, the type of pH buffer set, the pH Resolu
tion and enable/disable calibration Information display. To
enter Setup mode press and hold button until “STD
and tag is replaced by “SETUP“ and tag.
Release the button.
• “TEMP“ is displayed on the third
LCD line with the current temperature
unit (E.g. “TEMP °C“), for °C/°F
selection, use the SET button. After
the temperature unit has been selected, press to confirm
and to enter the “A-OFF“ selection.
Use the SET button, to cycle through
the auto‑off choices: 8 minutes (“8”,
default value), 60 minutes (“60”) or
disabled (“---”). Press to confirm
and to enter the “BEEP“ selection.
To switch ON or OFF the beep tone,
press the SET button; press to
confirm and to enter the calibration
buffer selection “pH 7.01 BUFF“.
To change the type of calibration
buffer set, the meter will show the
current buffer set: “pH 7.01 BUFF
(for standard buffer set: 4.01/7.01/
10.01) or “pH 6.86 BUFF” (for
NIST buffer set: 4.01/6.86/ 9.18). Change the set with the
SET button. Press to confirm and to enter pH resolution
selection “RESOL“.
To change the pH measurement
resolution between “0.1“ and “0.01
use the SET button; then press
to confirm and to enter electrode
calibration information “INFO“ selection.
METER SETUP
Impex Produkter

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI991003 and is the answer not in the manual?

Hanna Instruments HI991003 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI991003
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals