CERTIFICATION
PERIODIC MAINTENANCE
• INSPECT the electrode for any scratches or cracks. If any
present, replace the electrode.
• Rinse off any salt deposits with water.
• Follow the STORAGE procedure above.
CLEANING PROCEDURE
• Soak in Hanna HI7061 General Cleaning Solution for
approximately 20 minutes. Rinse well and calibrate
before using.
IMPORTANT: After performing any of the cleaning
procedures rinse the electrode thoroughly with distilled
water. Soak the electrode in HI70300 Storage Solution
for a minimum of 30 minutes. Rinse with water. Calibrate
before using.
TROUBLESHOOTING
• If the wick junction appears darkened, it may be pulled
out a few mm and the soiled part cut off.
• TEST: Evaluate your electrode performance based on the
SENSOR CHECK procedure on page 13.