17
ACCESSORIES
Code and serial
number on the
other side
Temperature sensor
ORP sensor
Reference junction
pH bulb
Quick connector with
Rubber protection, and
1 meter cable
PREPARATION
Remove the protective cap. DO NOT BE ALARMED IF ANY
SALT DEPOSITS ARE PRESENT. This is normal with elec
trodes and they will disappear when rinsed with water.
Shake the electrode down as you would do with a
clinical thermometer to eliminate any air bubbles inside
the glass bulb.
If the bulb and/or junction are dry, soak the electrode in
HI70300 Storage Solution for a minimum of 30 minutes.
Rinse with water.
Calibrate before using.
STORAGE
To ensure a quick response, the glass bulb and the junction
should be kept moist and not allowed to dry.
Replace protective cap with a few drops of HI70300
Storage Solution.
Follow PREPARATION above before taking any measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED WATER
ELECTRODE MAINTENANCE