measurement is always performed when the oscilloscope is running or stopped. 
Press [MEASURE] on the front panel to enter the measurement interface, and press F3 to select the 
DVM to enter DVM setting interface. 
Press CH1 Enable, CH2 Enable to enable any channel or all channels of DVM. 
Press CH1 Type, CH2 Type to select the data type displayed by DVM. 
DC RMS: displays the root-mean-squre value of the acquired data. 
AC RMS: displays the root-mean-squre value of the acquired data with the DC component removed. 
DC: Display the DC value of the acquired data. 
 
The display in the middle of the DVM box is the corresponding ratio of the current measured voltage 
value to the range corresponding to the eight vertical grids on the screen with the selected volt/div 
(turn the volt/div knob).   
2.12. Acquire 
2.12.1. Run Control   
Press the [Run/Stop] or [Single] button on the front panel to run or stop the sampling system of the 
scope.   
When  the  [Run/Stop]  is  green,  the  oscilloscope  is  running,  that  is,  acquiring  data  when  trigger 
conditions  are  met.  To  stop  acquiring  data,  press  the  [Run/Stop]  button.  When  stopped,  the  last 
acquired waveform is displayed.   
When the [Run/Stop] button is red, data acquisition is stopped. Red "Stop" is displayed next to the 
trademark logo in the status line at the top of the display. To start acquiring data, press [Run/Stop].   
To  capture  and  display  a  single  acquisition  (whether  the  oscilloscope  is  running  or  stopped),  press 
[Single] button. The Single run control lets you view single-shot events without subsequent waveform 
data overwriting the d isplay.   
When [Single] is pressed, the display is cleared, the trigger mode is temporarily set to Normal (to keep 
the  oscilloscope  from  auto-triggering  immediately),  the  trigger  circuitry  is  armed,  the  Single  key  is 
illuminated, and the oscilloscope waits until a user defined trigger condition occurs before it displays a 
waveform. 
When the oscilloscope triggers, the single acquisition is displayed and the oscilloscope is stopped (the 
[Run/Stop] button is illuminated in red). Press [Single] button again to acquire another waveform.