EasyManua.ls Logo

haTmed Q70B - Sterilization Procedure Completed

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 17 -
Fig4.3Manualstop
4.6 Sterilizationprocedurecompleted
Whenthesterilizationprocessendsnormally,theprogramendinterfaceandprogramrunningsuccess
informationaredisplayedonthescreen.Iftheautomaticprintingfunctionissetandtheprinterisconnected,the
productwillautomaticallyprintthesterilizationrecord.Pleasepress toopenthedoor.
Carefulburns
Do not forcethedoortoopen.Failuretodosomayresultin
damage to the product and/or hot steam coming out of the
chamber.
Afterthesterilizationisfinished,waitforawhileuntilthesterilizer
isfullycooledbeforeopeningthedoor.
Donottouchanymetalsurface.
Donottouchthesterilizedloadeditems,chamberanddoorcovers
withbarehandstoavoidbeingburned.
Besuretousethetrayhandletoremovetheloadeditems.
Balancewhenremoving thetraytopreventthe contentsof the
trayfromslipping.
Damagedpackagingdevicesoritemsmaynotbeeffectivelysterilized,
andexposuretothese items mayendangerthe healthofthe operatoror
patient.
Ifdamagedpackagingiffound,pleaserepackagetheitem.
Waiting for 5 minutes after sterilization before removing the
loadeditems,whichisgoodfordryingthesterileitems.
The 11th edition (2017) of the correct handling method for
medical device cleaning and disinfection issued by the
internationalauthorityAKI,Germany.Itisrecommendedthatthe
wet bag be judged as having water drops remaining on the
packageorinthepackageafter15minutesofsterilization.
Physical parameters should be monitored for each sterilization
cycle,recordandcheckprocessvariables.Whenthesterilization
temperaturesetinthiscycleis121°C,theholdingtimeshouldbe
nolessthan15minutes.Whenthesterilizationtemperaturesetin

Table of Contents