- 29 -
Cleanthefilterat
the bottom of the
sterilizationchamber
Sterilize200times
Wrench Unscrewthefilterwithawrenchandwashit
withcleanwater.
Cleantrays Sterilize100times
Rinseitwithcleanwater,ifstained,cleanit
withtoothpaste
Sterilizationchamber Everyweek
Wipethesterilizationchamberwithaclean
andwetmedicalcottoncloth,thenwipethe
chamberwithacleananddrymedicalcotton
cloth
8.2.2 Regularmaintenance
WS310‐2016 and WS506‐2016 require a comprehensive
maintenance of the small steam sterilizer every year to ensure
goodequipmentworkingconditionandsterilizationeffect
Maintenanceandtakecareofthisdevicemustbeperformedbya
professionalauthorizedbythemanufacturer.
9 Faultsandcommonproblemhandling
9.1 Nodisplayonthescreen
FaultydescriptionThedisplaydoesnotlightupnormallywhenthesterilizeristurnedon.
Solution
Checkifthereispowerinthemainpoweroutlet
Checkifthepowerplugofthisproductispluggedin.
Thescreenstilldoesnotappear,pleasecontactprofessionaldistributorauthorizedbyHӓTmed.
9.2 Largeconsumptionofdistilledwater
FaultydescriptionDistilledwaterconsumptionismuchlargerthanthespecifiedamount