EasyManua.ls Logo

haTmed Q70B - Helix Test; System Setting

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 21 -
5.3 HelixTest
TheHelixtestisusedtoverifythesteampenetrationandsaturationforAtypehollowinstruments,suchas
dentalhandpieces.SelecttheHelixtestinthetestprograminterface(Figure5.1).
Testmethodsandtestresultsarejudgedinaccordancewiththe
information and requirements provided by the test package
manufacturer.
5.4 Drainageprogram
Thisproductallowsuserstoperformdrainageprocedureswhenneeded.Thisprocedureisusedtodrainwater
remaininginasmallsteamsterilizer.
Whenthewaterinthechamberisobviousafteropeningthedoor,
itisrecommendedtousethedrainageprogrambeforeloadingthe
itemforsterilization.Especiallywhentheprogramisabnormally
stopped,thissituationiseasytooccur.Pleaseopenthedoorfor
inspection.
Please carry out the drainage program before handling this
product.
6 Systemsetting
Enter the system settings interface (Figure 6.1) for time, date, language, warm‐up, automatic printing,
additionaldryingsettings,andsystemreset.
Fig6.1Systemsetting
6.1 Timesetting
Thetimesettingoperationisasfollows:
1–Sterilizationprogram
2–Testprogram
3–Systemsetting
4–Filerecord
Press to enter
systemsetting
Timesetting
Datesetting
Languagesetting
Warm‐upsetting
Press to enter
timesetting
press and tosetup
time.Press toswitchbetween
time,minute and second.After
setting press to exit time
setting.
Timesetting
102739

Table of Contents

Related product manuals