- 19 -
5 Functionaltestingandotherprocedures
5.1 Vacuumtest
Whenfirstinstalled,unusedforalongtime,moved,orperiodicallytested,vacuumtestisrecommendedto
detectleakageinthesystemofthissterilizer:
Fig5.1VacuumTest
Vacuumtestshouldbecarriedoutwhenthesystemiscooledand
thechamberisdry.
In order to successfully complete the vacuum test,please check
thatthedoorofsterilizerisclosedandthecoolingwatervalveis
openbeforestartingtheprogram.
1–Sterilizationprogram
2–Testprogram
3–Systemsetting
4–Filerecord
Vacuumtestprogram
B&Dtestprogram
HELIXtestprogram
press toselect
07/25/201709:23:58
Vacuumtestprogram
15Min
T:25.0P:0.00Bar
Doorstatus
press toenter
press tostartup
Vacuumtestprogram
P:0.00BarT:25.0
Doorstatus
00:00:00