2.
Installation
and Connections
Transport
Connection to the
Power Supply
Place of installation
Adjustable feet
Ovens of thermal safety class 0 (see nameplate) must
be installed so as to permit permanent attendance. They
may not be stacked or built into existing laboratory
furniture, cf. section "Thermal Protection" (Page 14).
Care should be taken not to lift the oven by the door or the
doorhandle.
The ovens come with a power supply cable and a shock-
proof plug (16 A).
For line fusing, a 16 A delay-action fuse is required.
Prior to connecting the unit to the line, make sure that the
power supply ratings are those stated on the nameplate.
The connection must be made in accordance with the
relevant local safety regulations.
Rated voltage, rated frequency, rated power and current
are listed under 'Technical Data" and on the nameplate.
The oven must be placed on a level surface.
Its position is locked in if the adjustable feet are turned.
This requires two wrenches (size 24).