Connection for
central monitoring
system (potential-
free contact)
Provision to connect customer-installed warning
systems (e.g. telephone system, building control
systems).
Function:
When the thermal safeguard (temperature limit cut-out,
temperature limit controller) comes into action, or a timer
switches the oven on or off, the appropriate message will
be transmitted to the customer-installed warning system.
The potential-free contacts (1 make and 1 break contact)
are designed for the following circuits:
* Power supply circuits:
max. 250 V A.C., 3 A
* SELV/SELV-E circuits
(cf. VDE-Regulation 0100/410):
25/50 V A.C., 4 A
60/120 VD.C, 0.5 A
see also connection diagram
r
WPP API/ID 388
50012336
-16-