3.3 Temperature
Control
Mechanical
temp,
controller
"Heratherm"
Models 6030 - 6200 only
Temperature
range:
40 - 250 °C
Setting:
Use either a coin or a screwdriver to release the locking
device of the temperature controller.
Note the marking ! "j^j
Select the desired working temperature with the control
knob.
Tighten the locking device again.
Note the marking I
+
(
The set working temperature, once reached, is kept
constant with the controller opening and closing the heating
circuit as required. The white pilot lamp indicates that the
heater is working.
If working space temperature and dial readout do not
match,
the setting must be corrected (see Appendix).