1 Use according to specification
The centrifuge serves solely for separating human blood or bone marrow into its liquid (blood plasma) and solid
(erythrocytes, leukocytes) components in cell separation containers manufactured by Stryker Orthobiologics with the
help of a specially designed rotor.
The centrifuge is only meant for this purpose.
Another use or one which goes beyond this, is considered to be non-intended. The company Andreas Hettich GmbH
& Co. KG is not liable for damage resulting from this.
Observing all information in the operating instructions and complying with the measures described therein is also a
part of the intended use.
2 Remaining risks
The device is built according to the state-of-the-art and the recognized safety regulations. If used and handled
improperly, there could be life-threatening danger to the user or third parties, or the device could be impaired or there
could be other property damage. The device is only to be used for its intended purpose and only when it is in safe
working condition.
Malfunctions which could affect safety must be corrected immediately.
3 Technical specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model EBA 20 C
Type 2002-01C
Mains voltage (± 10%) 100 - 120 V 1
Mains frequency 50 - 60 Hz
Connected load 100 VA
Current consumption 0.8 A
Max. capacity 2 x cell separation containers, each filled to capacity
(approximately 10 ml of blood or bone marrow)
Allowed density
1.2 kg/dm
3
Speed (RPM) 4000
Force (RCF) 1574
Kinetic energy 850 Nm
Obligatory inspection (BGR 500) no
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Altitude
Up to 2000 m above sea level
Ambient temperature
2°C to 40°C
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Excess-voltage category
(IEC 60364-4-443)
ΙΙ
Pollution degree
2
Device protection class
Ι
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference,
Interference immunity
FCC Class B
Noise level (dependent on rotor)
63 dB(A)
Dimensions
Width
230 mm
Depth
304 mm
Height
216 mm
Weight approx. 4 kg