EasyManuals Logo

Hettich EBA 20 C User Manual

Hettich EBA 20 C
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
9/16
7 Initial operation
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Substances can be discharged through the ventilation opening on the rear side of the
centrifuge. The device is to be placed in such a way that the air stream is not directed towards
people.
Do not place any object in front of the ventiduct.
Keep a ventilation area of 300 mm around the ventiduct.
Centrifuge with PE connector:
If required, connect the PE connector at the back of the device with an additional medical potential equalization
system.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the connection cable to a standard mains socket. For connection ratings refer to
Chapter "Technical specifications".
Turn on the mains switch. Switch position "Ι".
The last used centrifuge data will be displayed.
Open the lid.
Remove the transportation safety device in the centrifuge, see sheet ”Transportation safety device”.
8 Opening and closing the lid
8.1 Opening the lid
The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on “Emergency release”.
Press the key
and open the lid. The symbol " " (lid open) illuminates in the rotation indicator .
8.2 Closing the lid
Do not bang the lid shut.
Place the lid and press the lid grip down slightly. The symbol " " (lid closed) illuminates in the rotation indicator
.
9 Loading the rotor
Check the rotor for firm seating.
Load both rotor positions equally. During centrifugation of only one cell separation container, the opposite rotor
position must be loaded with the supplied counter-weight, and the cell separation container must be loaded with
the maximum filling quantity as specified by Stryker Orthobiologics.
The weight of the permitted filling quantity is specified on the rotor. This weight must not be exceeded.
Always fill the cell separation containers outside of the centrifuge. Observe the specifications of the manufacturer
of the cell separation containers.
The maximum filling amount of the cell separation containers as specified by the manufacturer must not be
exceeded.
To keep the weight differences within the cell separation containers as low as possible maintain an even filling
height within the containers.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich EBA 20 C and is the answer not in the manual?

Hettich EBA 20 C Specifications

General IconGeneral
BrandHettich
ModelEBA 20 C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals