EasyManua.ls Logo

Hettich UNIVERSAL 320 - Programming; Write Protection for Programs; Opening or Loading Programs; Entering or Changing Programs

Hettich UNIVERSAL 320
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AB1401en_SARev.: 04 / 11.2023
32 / 44
Software operation
7.2 Programming
7.2.1 Write protection for programs
The programs can be protected to prevent unintentional changes.
When the rotor is at a standstill, write protection can be enabled or disabled
as follows:
1. Press and hold the
[SELECT]
button.
🡆
‘SOUND/BELL’
is displayed after 8 seconds.
2.
Press the
[SELECT]
button.
🡆
‘LOCK’
is displayed.
3. Use
[Rotary knob]
to set
‘OFF’
or
‘ON’
.
OFF = The programs are not write-protected
ON = The programs are write-protected
4. Press the
[START/IMPULS]
button.
🡆 The setting is stored.
If ON is set:
‘*** lock ***’
is displayed brieîž…y.
If OFF is set:
‘*** ok ***’
is displayed brieîž…y.
7.2.2
Opening or loading programs
1. Use the
[SELECT]
button to select the
‘PROG RCL’
parameter.
2. Use the
[Rotary knob]
to set the desired program location.
3. Press the
[START/IMPULS]
button.
🡆
‘*** ok ***’
is displayed brieîž…y.
The centrifugation data of the desired program location is dis-
played
4. To check the parameters: Press the
[SELECT]
button several times.
5. To exit the parameter indicator: Press the
[OPEN/STOP]
button or do
not press any button for 8 seconds.
7.2.3
Entering or changing programs
1.
Retrieve program.
2.
If required: Press the
[RCF]
button to toggle between RPM and RCF
indicator (
‘> <’
).
3.
If required: Press the
[SELECT]
button to select the desired parameter
and set it with the
[Rotary knob]
.
The parameters t/min and t/sec must be set to 0 using the
[Rotary
knob]
to set continuous operation. Continuous operation is shown in
the indicator with
‘¥’
.
4.
Use the
[SELECT]
button to select the
‘PROG STO’
parameter.
5.
Use the
[Rotary knob]
to set the desired program location.
6.
Press the
[START/IMPULS]
button.
🡆 Settings are stored in the desired program location.
‘*** ok ***’
is displayed brieîž…y.
The settings are always stored in program location # if the
[START/
IMPULS]
button is pressed without the
‘PROG STO’
parameter
being selected.

Table of Contents