35 / 44
Rev.: 04 / 11.2023
AB1401en_SA
Cleaning and care
7.5.3 Audible signal
7.5.3.1 General
The audible signal sounds:
after a problem occurs in the 2s interval.
after completion of the centrifugation run and rotor standstill in the 30 s
interval.
Opening the lid or pressing any button stops the audible signal.
7.5.3.2 Setting an audible signal
1. Press and hold the
[SELECT]
button.
🡆
‘SOUND / BELL ON’
or
‘SOUND / BELL OFF’
is displayed after 8
seconds.
2. Use
[Rotary knob]
to set
‘OFF’
or
‘ON’
.
OFF = audible signal disabled
ON = audible signal enabled
3. Press the
[START/IMPULS]
button.
🡆 The setting is stored.
‘*** ok ***’
is displayed briey.
8
Cleaning and care
8.1
Overview table
Chap. Task to execute
if required
daily
weekly
Annually
Page
8 Cleaning and care 35
8.3 Cleaning 36
8.3 Cleaning the device X 36
8.3 Cleaning the biosafety systems X 37
8.3 Cleaning the accessories X 37
8.4 Disinfection 37
8.4 Disinfecting the device X 37
8.4 Disinfecting the accessories X 37
8.5 Maintenance 38
8.5 Greasing the rubber seal of the centrifuging
chamber
X 38
8.5 Greasing the rubber seal of the biosafety system X 38
8.5 Trunnion greasing X 38