WI_Delivery Assist Trng CSM - DIR 50011929 ver F_2020-12-21 7
8. Power-on test
8.1. Plug the mains cable into a wall outlet
8.2. Confirm power led on the CSM is amber or green
8.3. Press the power button on the CSM to power-on the device and confirm there are no POST (Power On Self-Test)
errors
8.3.1. If the CSM has been pre-configured, it will power up to the Home screen, or the login screen.
8.3.2. If the device has not been pre-configured, the language selection screen appears, select the appropriate
language and press Exit”.
8.4. Set Date & Time
8.4.1. In the menu, go to Settings, press Date & Time to set date & Time according the instructions on page 28 of
the DFU
10
. Date & Time settings may not be available if the device has been configured to set this
automatically.
9. Configuration
9.1. If the CSM has been pre-configured, it will power up to the Home screen or login screen - no actions are
required during the delivery
9.2. If the CSM has not been pre-configured, follow these steps - you will receive a USB drive from the local Sales
Representative that contains the configuration profile for the monitors.
9.2.1. With the CSM switched on, insert the USB-flash drive with the configuration file(s) into the USB-host ports
on the underside of the CSM, or underneath the work surface of the APM mobile stand
9.2.2. A message will automatically appear on-screen allowing you to select the configuration you would like to
load. Press on the correct configuration within the list and then press “Load”
9.2.3. A Confirm box appears telling you that this will overwrite the configuration currently on the device. Press
OK”.
9.2.4. A message stating “Configuration Upload Successful” will appear at the top of the screen on the CSM.
9.2.5. Remove the USB flash drive
10. Distribute to nursing floor
10.1. Place the device in the desired location
10.2. Plug the power cord into the mains wall outlet to charge the battery
10.3. Instruct nursing staff to leave it plugged in for a minimum of 2 hours to fully charge the battery - if the
monitor is equipped with a Braun thermometer it will need to be charged for a full 12 hours.
11. Dispose of waste packaging
11.1. Dispose of packaging and pallet(s) according to local regulations
12. In-Service
12.1. Confirm appointment for in-service has been made with customer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Connex Spot Monitor 7000 Series and is the answer not in the manual?

Hillrom Connex Spot Monitor 7000 Series Specifications

General IconGeneral
BrandHillrom
ModelConnex Spot Monitor 7000 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals