38 The MetaNeb® System Instructions for Use (174432 REV 10)
Technical Specification for Performance—Breathing Simulator
HI PRESSURE CHFO PULSES—AIRWAY PRESSURE READING
HI PRESSURE CPEP—AIRWAY PRESSURE READING
CHFO Results
CHFO High Rate 240 ± 60 bpm (300 bpm max)
CHFO Low Rate 70% of CHFO High Rate (100 bpm min)
Peak Pressure 40 cm H
2
O
Oscillating Amplitude 5 cm H
2
O minimum