42 The MetaNeb® System Instructions for Use (174432 REV 10)
Bench Performance Checks—Breathing Simulator
Adult Restrictive
Breathing Simulator Settings—Bench Check 3
Lung Resistance 30 cm H
2
O/l/s
Lung Compliance 25 mL/cm H
2
O
Breath Rate 15 bpm
Effort Amplitude 15 cm H
2
O
Effort Slope 18
Effort % Inhale 33
Effort Offset 0 cm H
2
O
Target Volume 650 mL
Flow Conversion Factor 1
Time between Samples 30.1 milliseconds
Number of Samples 400