Heart and Respiration Rate Monitoring System powered by EarlySense
88 Centrella® Smart+ Bed with Heart and Respiration Rate Monitoring System powered by EarlySense Instructions for Use
(193587 REV 10)
look at the HR and RR data over time, see “HR and RR Trends Screen” on page 95. The heart and
respiration rate are shown in several screen views, see below.
If you do not want the heart and respiration rate to show on the screen, see “Turn ON/OFF HR and RR
Display” on page 94.
ALARM, ALERTS, AND INFORMATION INDICATORS
There must be either AC or battery power to the bed for the indicators to operate.
Low Priority Alarm
If the HR and RR option is active, a two tone alarm (e - c) will sound every 3 seconds to indicate that a HR
or RR alarm is active and a priority nurse call is sent to the nurses station (for beds equipped and
connected to a Nurse Call System).
Low Priority Alarm Parameters (HR/RR Monitoring System)
Beep Alerts
A single beep will sound when an activity is successful.
A triple beep will sound to indicate that you should look at the touchscreen, attention is needed.
For a critical fault, a triple beep will sound every 10 seconds until the user acknowledges the error on the
touchscreen. Follow the on-screen instructions. If the critical fault is not resolved, the triple beep will
sound again after an hour. The cycle will continue until the issue is resolved.
A low volume repeating triple beep every 10 seconds will sound to remind you that the Bed Exit is
silenced and should be resumed before you leave the room.
A quadruple beep (four beeps) will sound to indicate a Bed Exit alert.
A continuous tone will sound to indicate a serious bed malfunction and that immediate attention is
needed.
Home Screen Status Screen Trends Screen
The Information Indicators provide the caregiver with audible indicators and visual indicators.
Low Heart Rate High Heart Rate
Time to alarm for change in HR 90 seconds 90 seconds
Low Respiration Rate High Respiration Rate
Time to alarm for change in RR 180 seconds 180 seconds