What to do if my Hillrom PST 300+ Medical Equipment operating table won't charge and the ready display doesn't illuminate?
- JjoanneharrisAug 5, 2025
If your Hillrom Medical Equipment operating table isn't charging and the ready display [a5] isn't lit, there are several potential causes. First, inspect the mains power cable for any visible damage and replace it if necessary. Then, check and replace the power input fuse. If those aren't the issue, the main switch, voltage selector circuit board, main transformer, or power supply unit could be defective and require checking and replacement. Also, verify the connecting cable to the main board; if it's defective, replace it.