Tabletop cables
4990851_002_00 – 2082262 – 2022-01-10 93
13.2 Tabletop cables
Removal 1. Prepare the operating table (Chapter 5).
2. Disconnect the power supply on the operating table
(Chapter 6.1).
3. Remove the pad plate from the seat section (Chapter 7.2).
Attention: Leave the pad plate on the back section attached
so that the adjustment of the struts is maintained.
4. Remove the column head cover (Chapter 9.3).
5. Switch on the operating table using the switch at the running
gear.
6. Shift the longitudinal slide of the tabletop to the zero
position.
7. Switch off the operating table using the switch at the running
gear.
8. Note the cable positions and cable fastenings. Remove the
cable fastenings along the cable. If necessary, open the
power chain and remove the earth connection.
9. Pull out both plugs of the cable.
10. Remove the cable.
Assembly 1. Restore the original cable routing. Route the cable and cable
fastenings and restore any earth connection. Close the
opened power chain.
2. Attention: Do not interchange the plug connections. Pay
attention to the wiring diagram on page 64.
Restore the original cable routing. Connect both plugs for the
cable.
3. Mount the column head cover (Chapter 9.3).
4. Mount the pad plate onto the strut (Chapter 7.2).
5. Put the pad in place.
6. Connect the power supply to the operating table
(Chapter 6.2).