Column cables
4990851_002_00 – 2082262 – 2022-01-10 103
14.2 Column cables
Removal 1. Prepare the operating table (Chapter 5).
2. Disconnect the power supply on the operating table
(Chapter 6.1).
3. Open the column top and guide the column cover
downwards (Chapter 6.5).
4. Note the cable positions and cable fastenings. Remove the
cable fastenings along the cable. If necessary, open the
power chain and remove the earth connection.
5. Pull out both plugs of the cable.
6. Remove the cable.
Assembly 1. Restore the original cable routing. Route the cable and cable
fastenings and restore any earth connection. Close the
opened power chain.
2. Attention: Do not interchange the plug connections. Pay
attention to the wiring diagram on page 64.
Restore the original cable routing. Connect both plugs for the
cable.
3. Guide the column cover upwards and close (Chapter 6.6).
4. Put the pad in place.
5. Connect the power supply to the operating table
(Chapter 6.2).