Maintenance
110
4990851_002_00 – 2082262 – 2022-01-10
2. Visual inspection of adhesive labels for any damage and
whether they are easy to read
3. Verification of column covers
4. Verification of bellows
5. Verification of all seals and cogged belts in the operating
table
6. Verification of all mechanical coupling elements
19.1 List of parts subject to wear
1All wearing parts are to be changed as part of the regular
maintenance work and if otherwise necessary.
With proper use, care and maintenance of the operating table, the
following parts are declared to be high-wear parts:
19.2 Lubrication schedule
Consumable Replace if faulty
Tabletop
All side rails •
All cogged belts •
All pads •
Column
Bellows •
All cogged belts •
Running gear
Batteries •
Jack-up unit foot plate •
Fuses •
Mains power cable •
Lubrication points Lubricant Application Interval to
relubricate
Tooth rod, longitudinal
slide
ISOFLEX TOPAS NCA 52 with brush 2 years
Trendelenburg spindle ISOFLEX TOPAS NCA 52 with brush 2 years
Edging spindle ISOFLEX TOPAS NCA 52 with brush 2 years
Lift spindle POLYLUB GLY 151 with brush 2 years
Trendelenburg bearing ISOFLEX TOPAS NCA 52 with grease gun 2 years
Wheel suspension spring UNIMOLY GB 2
(#2065940)
with brush 2 years
Contact surface between
guide and sleeve on the
emergency release
UNIMOLY GB 2
(#2065940)
with brush 2 years