Running gear mechanical parts
4990851_002_00 – 2082262 – 2022-01-10 79
11.8 Locking bolt/locking bolt hinge jack-up unit
Removal 1. Prepare the operating table (Chapter 5).
2. Release the parking brake of the operating table.
3. Disconnect the power supply on the operating table
(Chapter 6.1).
4. Open the column cover on the running gear (Chapter 6.3).
5. Loosen the running gear metal cover [1] (4 screws [2]).
6. Push up the running gear metal cover [1] and fix in the upper
position.
7. Remove the emergency release lever (Chapter 11.6).
8. Remove the deflector plate [4] at the top of running gear (6
screws [3]).
9. Remove the jack-up unit linear drive (Chapter 12.2).
10. Remove the emergency release for jack-up unit [5]
(Chapter 11.7).
11. Remove the locking bolt/locking bolt hinge [6] from the
running gear (4 screws [7]).
Assembly 1. Mount the locking bolt/locking bolt hinge on the running gear
(4 screws).
2. Mount the emergency release for jack-up unit (Chapter 11.7).
3. Mount the deflector plate at the top of running gear (6
screws).
4. Mount the jack-up unit linear drive (Chapter 12.2).
5. Mount the emergency release lever (4 screws).
6. Mount the running gear metal cover on the running gear
(Chapter 11.2).
7. Guide the column cover downwards and close on the running
gear (Chapter 6.4).
8. Apply the parking brake of the operating table.
9. Put the pad in place.
10. Connect the power supply to the operating table
(Chapter 6.2).