Safety
14
4990851_002_00 – 2082262 – 2022-01-10
1.9 Components with serial number
Individual, safety-relevant components or assemblies of the
operating table are marked with an additional serial number.
When replacing one of the listed parts, the serial numbers of the
removed and new parts must be reported to Trumpf
Medizin Systeme using the “Service Report” form (country-
specific). Complete the “Service Report” form in full and fax or
email it to the Technical Customer Service. This information is
required in order to process subsequent customer requests
correctly. Attention - it may be the case that assemblies contain
several components with serial numbers. In this case, all
corresponding serial numbers must be entered in the “Service
Report” form.
The “Service Report” form can be downloaded from the Trumpf
Medizin Systeme online information system. On completion of
work, package the defective part securely and send it back to
Trumpf Medizin Systeme.
The following components or assemblies on the operating table
can be clearly identified by their serial numbers:
Tabletop:
Pos. Material description Material
number
1 Motor for back section hinge 1977822
2 Motor for leg section hinge 1977822
3 Longitudinal slide motor 1977822
4 PCBA distributor board 2081702
5 Connector socket for the control module
with cable connection
2023321
6 Connector socket for the control module 2023322
1
2
2
1
3
4
6
5