Running gear electric components
4990851_002_00 – 2082262 – 2022-01-10 91
12.12 Battery
Removal 1. Prepare the operating table (Chapter 5).
2. Disconnect the power supply on the operating table
(Chapter 6.1).
3. Open the column cover on the running gear (Chapter 6.3).
4. Loosen the running gear metal cover [1] (4 screws [2]).
5. Push up the running gear metal cover [1] and fix in the upper
position.
6. Remove both battery fuses (Chapter 12.11).
7. Note the connection assignments. Pull out two connecting
cables for each battery [7] (on the battery set).
8. Remove the deflector plate [4] at the foot end of the running
gear (2 screws [3] with washer).
9. Remove both battery attachment frames [6] (2 screws [5]
each).
10. Note the mounting position of the battery set. Remove the
complete battery set.
Assembly 1. Return the batteries to their original mounting position. Insert
the new battery set and position correctly.
2. Mount both battery attachment frames (2 screws each).
3. Mount the deflector plate at the foot end of the running gear
(2 screws with washer).
4. Attention: do not interchange the battery connections. Pay
attention to the wiring diagram on page 64.
Restore the original connections. Connect two connection
cables to each battery. The voltage between brown and
white is 24 VDC.
5. Insert both battery fuses (Chapter 12.11).
6. Mount the running gear metal cover on the running gear
(Chapter 11.2).
7. Guide the column cover downwards and close on the running
gear (Chapter 6.4).
8. Put the pad in place.
9. Connect the power supply to the operating table
(Chapter 6.2).
2
1
5
7
4
6
7
3

Table of Contents