Preventive Maintenance
130 Centrella® Smart+ Bed Instructions for Use (193587 REV 10)
The zipper tabs are located on the left side at the head end of the mattress.
b. Machine wash the top cover per your facility protocol. The cover can be washed at a maximum
water temperature of 54°C (130°F).
• Use a disinfectant as instructed in the manufacturer’s instructions.
• To determine the amount of disinfectant to use, determine the amount of water in the washer,
and follow the manufacturer’s dilution instructions.
• During the wash cycle, soak the top cover in the disinfectant.
• Let the top cover rinse thoroughly in clean water.
c. Use the lowest temperature setting of the dryer to dry the top cover; do not exceed 43°C
(110°F).
d. Follow the Cleaning and Disinfecting instructions. See “Cleaning and Disinfecting” on page 124.
For rental beds, Hill-Rom Service personnel will follow the Laundry Wash and Dry Procedure (QS02040).
PREVENTIVE MAINTENANCE
Warning—Only facility-authorized persons should service the Centrella® Smart+ Bed. Service done by
unauthorized persons could cause injury or equipment damage.
The Centrella® Smart+ Bed requires an effective maintenance program. We recommend that you do
annual preventive maintenance (PM) for Joint Commission certification. PM not only meets Joint
Commission requirements but can help make sure of a long, operative life for the Centrella® Smart+ Bed.
PM will help minimize downtime due to excess wear. For a preventive maintenance schedule, refer to
the Centrella® Smart+ Bed Service Manual (193588).
If the bed has the pro+ non-integrated mattress (P7924A), see the pro+ Mattress Service Manual (209197)
for the mattress preventative maintenance.
For service and/or technical information other than that specified in this manual, including fuse
replacement, circuit diagrams, and isolation of AC power, refer to the Centrella® Smart+ Bed Service
Manual (193588).