Chapter 6 Maintenance and Service
___________________________________________
Trouble Cause Action
Nothing is
displayed when
power is turned
on.
Is a battery
installed?
See '3.1:
Preparations for the
Batteries.'
Is the battery
dead?
Replace with a fresh
battery.
No temperature
measurement is
displayed. (O.F
or ----- is
displayed)
Is the temperature
probe connected
correctly?
Reconnect the
temperature probe.
Is the sensor
disconnected?
Replace with a new
sensor.
E92 is displayed.
E97 is displayed.
Built-in memory
damaged due to
some reason.
Requires repair.
(Read "Service"
below.)
6.2 Troubleshooting (before calling for
hel
Service
If the instrument seems to be malfunctioning,
confirm that the batteries are not discharged, and
that the temperature probes are not open circuited
before contacting your dealer or Hioki
representative.
When sending the instrument for repair, remove
the batteries and pack carefully to prevent damage
in transit. Include cushioning material so the
instrument cannot move within the package. Be
sure to include details of the problem. Hioki
cannot be responsible for damage that occurs
during shipment.