5
___________________________________________
Chapter 2 Specification
___________________________________________
Measurement Sensor Thermocouple, K type
Measurement Range -100 to 1300 , -148 to 2372
Resolution 0.1 [-100.0 to 199.9 ]
1 [200 to 1300 ]
0.1
[-148.0 to 391.9 ]
1 [392 to 2372 ]
AUTO RANGE
Accuracy
(23
5 )
(73 9 )
Period of Guaranteed Accuracy:
6months
(0.1% rdg.+0.8 ) [-100.0 to 199.9 ]
(0.1% rdg.+1.4 ) [-148.0 to 391.9 ]
(0.2% rdg.+1 ) [200 to 1300 ]
(0.2% rdg.+1.8 ) [392 to 2372 ]
Period of Guaranteed Accuracy:
1 year
(0.15% rdg.+1.2 ) [-100.0 to 199.9 ]
(0.15% rdg.+2.1 ) [-148.0 to 391.9 ]
(0.3% rdg.+1.5 ) [200 to 1300 ]
(0.3% rdg.+2.7 ) [392 to 2372 ]
Temperature
Coefficient
0.03 / [-100 to 199.9 ]
0.05 / [200 to 1300 ]
0.05
/ [-148.0 to 391.9 ]
0.09 / [392 to 2372 ]
Sampling Time 0.5s