Measure the commercial power supply
9
Data will be saved under the pre-set saving conditions when the SAVE key is pressed.
Refer to "Analyze"( p.12) for the analyzing method.
3 Set the saving conditions
Move
the cursor
Displays the File
Save sheet.
The following explains the setting method to save immediately by pressing the SAVE key.
Carry out the following setting using the File Save sheet on the System screen.
Select setting
items
*: When [Yes] is set, pressing the SAVE key each time will allow users to select
the saving conditions.
Point
To view the waveform on the instrument, set the [Save Type] to binary and to view the waveform
on the computer, set it to text. Data saved as text cannot be viewed on the instrument.
Select at save (Selection Save)
*
Set to [No].
Pressing the SAVE key will allow sav-
ing under the saving conditions to be
carried out immediately.
Save conditions
Save Type: Wave Binary
Save to: CF Card
Save Name: TEST
Save Area: Whole Wave
The setting will appear as shown
on the left screen.
4 Start to stop the measurement
Press the START key.
When the trigger conditions are met, the trigger is
activated and the waveforms equivalent to the pre-
set record length will be recorded.
Trigger level
0 V
Trigger slope [ ]
T
Measurement data will be recorded until the
STOP key is pressed.
[Trigger Wait]
5 Save data
Point
Saved data can be confirmed on the File screen when the FILE key is pressed. ( p.16)