EasyManua.ls Logo

Hioki 8847 - 1.4 Basic Key Operations

Hioki 8847
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Others
22
Set the length (number of divisions) to record each time data is acquired. Each division of the recording
length consists of 100 data points. The number of data points of the whole recording length is the specified
recording length (divisions) × 100, + 1.
Example. The number of data points when the specified recording length is 50 divisions:
50 div × 100 data points + 1 = 5001 data points
Optical resolution differs with different input units.
The following table shows the full scale optical resolution for all the units.
The smallest optical resolution can be calculated from the screen full scale value and the full scale optical
resolution shown in the table below.
E.g. In the case of measurement with the 8966 Analog unit
The smallest optical resolution when the power voltage with vertical axis 20V/div and vertical axis mag-
nification x 1 is measured is
Screen full scale: 20 V/div x 20 div = 400 V
Full scale optical resolution at vertical axis x 1: 2000
400 V / 2000 = 0.2 V
Brackets indicate valid data range
*: With the 8967 Temp Unit, the valid range differs depending on the thermocouple. For information on the minimum resolution,
see the specifications of the 8967 Temp Unit.
The following shows the speeds of saving binary data (reference values) using different media and inter-
faces. Note that the data saving speed varies depending on the saving conditions, device manufacturer,
device capacity, communication conditions, and others.
Recording Length Setting
Voltage axis and optical resolution
Full-scale resolution for input units at various vertical axis zoom factors (LSB)
Input module Zoom factor
×1/10 ×1/5 ×1/2 ×1 ×2 ×5 ×10 ×20 ×50 ×100
8966 (Analog)
8971 (Current)
8972 (DC/RMS)
20000
(4000)
10000
(4000)
4000 2000 1000 400 200 100 40 20
8967
(Temperature)
*
200000 10000 40000 20000 10000 4000 2000 1000 400 200
8968
(High resolution)
320000
(64000)
160000
(64000)
64000 32000 16000 6400 3200 1600 640 320
8969 (Strain) 250000
(64000)
125000
(64000)
50000 25000 12500 5000 2500 1250 500 250
8970
(Power frequency )
20000 10000 4000 2000 1000 400 200 100 40 20
8970 (Count ) 400000 200000 80000 40000 20000 8000 4000 2000 800 400
8970 (Excluding
power frequency
and count)
100000 50000 20000 10000 5000 2000 1000 500 200 100
Data Saving Speed
Storage media Saving speed (reference value)
Save to CF card 500kB/s
Save to HDD 800kB/s
Save to USB 500kB/s
Save to PC with LAN 500kB/s