Analyze
15
Up to 16 can be calculated at once.
Calculation items: average value, execution value, peak value, maximum value, minimum value, cycle, fre-
quency, etc. The calculation method on the measurement data for all the 20 items will be explained here.
When setting window, etc. is displayed in the waveform screen, pressing the ESC key will display the cal-
culation results.
Calculate the measurement data
2. Set the
calculation type
3. Select the calculation
target channel
4. Set the parameters
(calculation conditions)
1. Set to [On].
Move
the cursor
Displays the
[Num Calc] sheet
Carry out the following setting using the calculation sheet on the status screen.
Select
setting
items
Once measurement is completed, calculation automatically starts. The calculation result will be dis-
played on the top right side of the waveform screen.
Displays the
Waveform screen.
In consecutive trigger mode, next measurement will take
place after calculation has been executed.