Measurement Preparations
5
Supply the power
Power On
Right Side
Setting the clock
Carry out setting in the [Init] sheet. The date and time indication is shown
at the top right of the screen.
Perform zero-adjust
This procedure compensates for input module differences and sets the reference
potential of the instrument to 0 V.
Carry out setting in the [Unit List] sheet.
For best measurement precision, a warm-up period of about 30 minutes after power-on is recommended.
[Exec Zero-Adjust] can still be selected in the channel settings window ([Analog]
sheet).
(Press the DISP key (Waveform screen) press the CH.SET key (Channel settings
window))
Point
1. Move the cursor to this position.
2. Select [Exec Zero-Adjust].