Chapter 3 Specifications
General Specifications
Guaranteed
accuracy period
1 year
Operating
Environment
Indoors, Pollution Degree 2
Altitude up to 2000 m (6562-ft.)
Operating
Temperature &
Humidity
0 to 40
C (32 - 104
F)
90%RH or lower (non-condensating)
40 to 50C
(104 - 122
F)
, at 50C and below relative
with linear decrease up to 50%
RH
Storage Tempera-
ture &Humidity
-10 to 50°C (14 - 122°F)
90%RH or lower (non-condensating)
Degree of
protection
IP40
Maximum rated
voltage to
terminal
600 V AC (150 V AC for model IR4030)
AC voltage function
Maximum rated
voltage to earth
600 V AC, Measurement Category III,
Anticipated Transient Overvoltage: 6000 V
(2500 V AC for model IR4030)
Dielectric strength
7060 V AC (2700 V AC for model IR4030),
50/60 Hz, Measurement terminals - electrical enclo-
sure, 1 minute, current sensitivity 1 mA
Power source
Rated power voltage: 1.5 V DC × 4
LR6 alkaline battery × 4
Maximum rated
power
3 VA
Drop Proof On concrete: 1 m
Dimensions
Approx. 152W
×
92H
×
40D mm
(5.98"W×3.62"H×1.57"D)
(not including protrusion)
Mass
Approx. 420g
(14.8
oz
)
(including battery, not including test lead)
Options
L9787 Test Lead, L9787-91 Breaker Pin (for Model
L9787), L9788-10 Test Lead with Remote Switch,
L9788-11 Test Lead Set with Remote Switch, L9788-
90 Tip Pin (for Model L9788-10), L9788-92 Breaker
Pin (for Model L9788-10), 9804-02 Magnetic Adapter,
C0100 Carrying case
Standards
EMC EN61326
Safety EN61010