Chapter 3 Specifications
索引
4
3
2
1
7
6
5
10
9
8
Accessories
*IR4082 has -11 only.
Common Specifications
Model
Name
IR4011, IR4012,IR4013, IR4014, IR4015, IR4030,
IR4031, IR4032, IR4033, IR4041, IR4042, IR4082*
Version -10 ‐ 11
Accessories
L9787 Test Lead×1
L9788-11
*
Complete Test
Leads with Remote Control
Switch ×1
C0100 Carrying case ×1
Instruction manual ×1
LR6 alkaline battery × 4
Accuracy (allowance) guarantee range
• Ambient temperature
• Relative humidity
• Position
• Magnetic field
• Supply voltage
23°C±5°C (73
F±9
F)
90%
RH or lower (with no condensations)
Horizontal ±5
Non (Earth’s magnetic field)
Battery effective range
Voltage Measurement
• Measuring range
• Accuracy
• Frequency range
• Effect of temperature
• Effect of position
• Input resistance
0 to 600 V (IR4030: 0 to 150 V)
±5% of maximum indicated value
50 Hz/60 Hz
±5% of maximum indicated value
±2% of maximum indicated value
500 k or more (50 Hz/60 Hz)
(100 k or more for model IR4030)
Insulation Resistance Measurement
• 1st effective measur-
ing range
(Intrinsic error)
±5% of indicated value
• 2nd effective measur-
ing range
(Intrinsic error)
±10% of indicated value
• 0 Mscale 0.7% of the scale length
•Operating
1st effective measuring range:±25% of indicated value.