Chapter 3 Specifications
34
Model IR4011, IR4012, IR4013, IR4014, IR4015
Rated measurement
voltage (DC)
125 V 250 V 500 V 500 V 1000 V
Maximum effective
indicated value
100 M 1000 M 2000 M
Center scale value
2 M 20 M 50 M
1st effective
measuring range
0.1 M to 50 M
1 M to
500 M
2 M to
1000 M
2nd effective
measuring range
0.01 M to 0.1 M
50 M to 100 M
0.5 M to
1 M
500 M
to 1000
M
1 M to
2M
1000M
to
2000M
Lower limit measure-
ment
resistance value to be
maintained rated
measurement voltage
0.125
M
0.25
M
0.5 M
0.5 M 1 M
Continuous operating
time
Approx. 20 hours (no load) 15 hours
Input error protection for
10 seconds
600 VAC 660 VAC