Chapter 3 Specifications
索引
4
3
2
1
7
6
5
10
9
8
Model
IR4082
Insulation Resistance measurement
Rated measurement
voltage (DC)
125 V 250 V 500 V
Maximum effective
indicated value
100 M
Center scale value
2 M
1st effective
measuring range
0.1 M to 50 M
2nd effective
measuring range
0.01 M to 0.1 M
50 M to 100 M
Lower limit measurement
resistance value to be
maintained rated
measurement voltage
0.125 M 0.25 M 0.5 M
Continuous operating
time
Approx. 15 hours (Measured when measurement
terminal intervals are opened in the 500 V Range)
AC resistance measurement
Measurement Range
×1 ×10
Display Value
Effective Resistance Value
Maximum effective
indicated value
110 1100
Measuring range
2 to 110 20 to 1100
Accuracy
±
3
±
30
Graduation Interval
2
20
Reactance Allowance
Range
12
or less (Distance to common earth used: approx. 500 m)
At Measurement Frequency
Open Circuit Voltage
38 Vpk
Measurement Current
2.36 mArms
±
0.13 mArms
Frequency range
575 Hz
±
10 Hz
Measurement Current
Waveform
Sine wave
Effect of temperature
±
3
±
30
Effect of position
±
3
±
30
Effective Earth Voltage
14.2 Vpk
Effect of Earth Voltage
0.5
Vrms
or less
Within product specification
0.
5 to 10
Vrms
±
3
±
30
(for DC or 50/60 Hz sine waves)
Effect of Voltage (in the
Battery Effective Range)
±1.5 and within
specification allowance
±15and within
specification allowance
Magnetic field
interference
(400A/m)
±
1.5
±
15
Measurement Range
×1 ×10