Chapter 3 Specifications
36
Effect of Earth Capacity
10 nF or less
within specifications
within specifications
10 nF - 20 nF
±
24
20 nF - 30 nF
±
48
30 nF - 40 nF
±
72
40 nF - 50 nF
±
120
50 nF
-
100 nF
Not specified
100 nF
-
200 nF
±
1.2
200 nF
-
300 nF
±
2.4
300 nF
-
400 nF
±
4.8
400 nF
-
500 nF
±
7.2
500 nF or more
Not specified
Input error protection
Insulation Resistance
Range
600 VAC
(
10 seconds
)
AC Voltage Range
600 VAC
AC Resistance
Range
50/60 Hz 600 V (10 seconds)
Earth Voltage
Check
50/60 Hz 600 V (10 seconds)