8.6 Message Reference
154
Measurement Flow
Any of the fol-
lowing:
TRIG
Termi-
nal
TRIG Key
Any of the fol-
lowing:
TRIG
Termi-
nal
TRIG Key
TRG
:INITIATE:CONTINUOUS ON
:TRIGGER:SOURCE IMMEDIATE
Measurement
Calculation
Measured Value Output
Idle State
Trigger Delay
Measured Value Output
:INITIATE:CONTINUOUS OFF
:TRIGGER:SOURCE IMMEDIATE
Measurement
Calculation
:INITIATE:IMMEDIATE
Trigger Wait State
Trigger Delay
Measured Value Output
Measurement
Calculation
:INITIATE:CONTINUOUS ON
:TRIGGER:SOURCE EXTERNAL
Idle State
Trigger Delay
Measured Value Output
Measurement
Calculation
Trigger Wait State
:INITIATE:CONTINUOUS OFF
:TRIGGER:SOURCE EXTERNAL
:INITIATE:IMMEDIATE
11 2
3 4
Trigger Delay
HIOKI BT3562A981-09