EasyManua.ls Logo

Hioki BT3562 - Sample Programs; To be Prepared in Visual Basic 5.0;6.0

Hioki BT3562
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.8 Sample Programs
166
These sample programs are written in Microsoft Visual Basic
®
5.0 and 6.0.
The following are used for communication:
For RS-232C communication: MSComm from Visual Basic
®
Professional
For GP-IB communication: National Instruments GP-IB Board, Driver and
Module for Visual Basic
®
During communications, the terminator setting is supposed to be as follows:
RS-232C: CR+LF
GP-IB: LF
RS-232C Communications_______________________________________
(Using Microsoft Visual Basic
®
Professional MSComm)
(1) Simple Resistance Measurement
Imports measured values 10 times, and saves measurements in a text file.
8.8 Sample Programs
To be prepared in Visual Basic
®
5.0/6.0
Private Sub MeasureSubRS()
Dim recvstr As String 'Receiving char string
Dim i As Integer
MSComm1.Settings = "9600,n,8,1" 'Comm port setting
MSComm1.PortOpen = True 'Open a port
Open App.Path & "\data.csv" For Output As #1 'Open a text file for saving
MSComm1.Output = ":TRIG:SOUR IMM" & vbCrLf 'Select internal triggering
MSComm1.Output = ":INIT:CONT ON" & vbCrLf 'Continuous measurement ON
For i = 1 To 10
MSComm1.Output = ":FETCH?" & vbCrLf 'Send ":FETCH?" to import the most recent mea-
surement
recvstr = "" 'From here on, continue receiving until an LF code
occurs
While Right(recvstr, 1) <> Chr(10)
recvstr = recvstr + MSComm1.Input
DoEvents
Wend
recvstr = Left(recvstr, Len(recvstr) - 2) 'Delete the terminator (CR+LF)
Print #1, Str(i) & "," & recvstr 'Write to the file
Next
Close #1
MSComm1.PortOpen = False
End Sub
HIOKI BT3562A981-09