4.3 Measurement Current Pulse Output Function
61
4
Chapter 4 Applied Measurement
When using multiple instruments to take measurements at the same time, you
may experience interference as a result of electromagnetic induction caused by
the devices' measurement currents, causing measured values to gradually oscil-
late between increasingly large values. To prevent this phenomenon, the mea-
surement current pulse output function can be used to apply the measurement
current only while measurement is being performed. (This functionality is only
available during external trigger measurement or while continuous measurement
is off.)
When this function is turned on, the measurement current is only applied at the
time of measurement. By staggering the timing of trigger measurement, mea-
surement current interference can be eliminated.
4.3 Measurement Current Pulse Output Function
1
(The SHIFT indicator lights up.)
The menu screen is displayed.
2
Display the measurement current pulse function configuration screen.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16)
3
Set the constant current pulse function to either "On" or "Off."
on.................... Measurement current pulse function on.
oFF.................. Measurement current pulse function off.
4
Accept the setting and return to the measurement screen.
HIOKI BT3562A981-09